Correct use of multi-purpose vacuum pump for circulating water

the body of multi-purpose vacuum pump for circulating water adopts double taps, which can be used alone or in parallel. Two vacuum meters are installed. The main machine is made of stainless steel movement and anti-corrosion material movement. It is corrosion-resistant, pollution-free, low noise and convenient to move. It can also be equipped with a vacuum regulating valve according to the needs of users. Four students can conduct chemical experiments at the same time to reduce the experimental space. It has the characteristics of small volume, light weight and beautiful appearance. It has double tables, double heads for air extraction, four tables and four taps, and a multi-purpose vacuum pump with the same two sides. Even if the teacher visually demonstrates, students can turn on and off on any side. The product is specially equipped with double-layer glass reaction kettle, rotary evaporator, vacuum drying oven, rotary film evaporator, vacuum suction filter, filter bottle, multi-functional reactor, vacuum film concentrator and other best products. It is one of the indispensable ideal products in the laboratory

the correct use of multi-purpose vacuum pump for circulating water is as follows:

1. Regularly check the wear of the shaft sleeve, and replace it in time after the wear is large. When the vacuum pump is used in cold winter, after stopping, unscrew the drain plug at the lower part of the pump body to drain the media. Prevent frost cracking. If the vacuum pump is out of service for a long time, it needs to be amplified and converted by the a/d converter. Disassemble the pump completely, wipe off the moisture, coat the rotating parts and joints with grease, install them, and keep them properly

2. When it takes a long time to work, the water temperature in the water tank will rise, which will affect the vacuum degree. At this time, connect the drain hose with the water source (tap water), and the overflow nozzle will be used as the drainage outlet. Properly control the flow of tap water, so as to maintain the water temperature in the water tank and maintain the stability of the vacuum degree

3. If the vacuum degree does not go up, first judge whether it is leaked by the pumping container or the leather pipe is loose. If it is a pump problem, check whether the water inlet or each air circuit is blocked or loose, and whether there is air leakage. If the motor of the circulating water vacuum pump does not rotate, check the power supply or fuse

4. Generally, the vacuum pump should be overhauled after 2000h of operation, check the aging degree of rubber seals, check whether the exhaust valve disc is cracked, and clean up the dirt deposited on the valve disc and exhaust valve seat. Clean the parts in the whole vacuum pump cavity, such as rotors, rotating blades, springs, etc. It is usually washed with gasoline and dried. After cleaning the rubber parts, dry them with a dry cloth. When cleaning and assembling, handle with care to avoid bumping. Clean the pipe if possible to ensure that the pipeline is unblocked. Check whether the vacuum pump pipeline and joint are loose

5. When it is necessary to provide cooling circulating water for the reaction device, on the basis of the previous three operations that can reduce (increase) the temperature of the experimental container outside the tank, connect the water inlet and outlet pipes of the device to be cooled to the circulating water outlet and water inlet nozzles at the rear of the machine respectively, and turn the circulating water switch to the on position to realize the supply of circulating cooling water

6. The multi-purpose vacuum pump for circulating water should be commissioned after reassembly. Generally, it must run idle for 2h and change the oil twice. Because a certain amount of volatile substances will be left in the vacuum pump during cleaning, it can be put into normal operation after normal operation. Add bearing lubricating oil into the bearing body, observe that the oil level should be at the center line of the oil standard, and replace or supplement the lubricating nano friction and wear test oil in time. Unscrew the water diversion plug of the vacuum pump body and fill the water diversion (or slurry diversion). Close the gate valve, outlet pressure gauge and inlet vacuum gauge of the water outlet pipeline

7. Keeping the water clean is the key to the long-term stable operation of the equipment. The water tank must be changed and cleaned regularly. Do not pump dust and solid substances

8. Turn the vacuum pump by hand to see if the vacuum pump is flexible. Jog the motor to see if the motor turns correctly. Start the motor. When the vacuum pump operates normally, turn on the outlet pressure gauge and the inlet vacuum pump. After it shows appropriate pressure, gradually open the gate valve and check the motor load at the same time. Try to control the flow and head of the circulating water vacuum pump on the label, and then provide a more comprehensive performance better than ABS in most cases within the specified range, so as to ensure that the vacuum pump operates at the highest efficiency point, so as to obtain the maximum energy-saving effect. During the operation of the vacuum pump, the bearing temperature shall not exceed the ambient temperature by 35 ℃, and the maximum temperature shall not exceed 80 ℃

9. Some corrosive gases can cause the water quality in the water tank to deteriorate, produce bubbles, and affect the vacuum degree. Therefore, attention should be paid to continuous circulation and water change. For special highly corrosive gases, carefully judge whether they react with the materials used in this equipment, and use them carefully

10. Check the oil level frequently. If it does not meet the requirements, adjust it to meet the requirements. When the vacuum pump is running, the oil level shall be subject to the center of the oil standard. Check the oil quality frequently. If the oil is deteriorated, replace it with new oil in time to ensure that the vacuum pump works normally

11. If abnormal sound is found in the multi-purpose vacuum pump of circulating water, stop immediately to check the cause. When the vacuum pump is to be stopped, close the gate valve and pressure gauge first, and then stop the motor. Within the first month of operation, the lubricating oil of the vacuum pump shall be changed after 100h, and then every 500h. Adjust the packing gland frequently to ensure that the leakage in the packing chamber is normal (it is appropriate to leak in drops)